• Users Online: 232
  • Print this page
  • Email this page
ORIGINAL ARTICLE
Year : 2021  |  Volume : 33  |  Issue : 1  |  Page : 42-48

Effects of different electrodes used in bone-guided extracochlear implants on electrical stimulation of auditory nerves in guinea pigs


1 Department of Mechanical Engineering, National Taiwan University, Taipei, Taiwan
2 Department of Materials and Mineral Resource Engineering, Institute of Materials Science and Engineering, National Taipei University of Technology, Taipei, Taiwan
3 Department of Otolaryngology, Hualian Tzu Chi Hospital, Buddhist Tzu Chi Medical Foundation; Department of Otolaryngology Head and Neck Surgery, School of Medicine, Tzu Chi University, Hualien, Taiwan

Correspondence Address:
Dr. Chia-Fone Lee
Department of Otolaryngology, Hualian Tzu Chi Hospital, Buddhist Tzu Chi Medical Foundation, 707, Section 3, Chung-Yang Road, Hualien
Taiwan
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/tcmj.tcmj_46_20

Rights and Permissions

Objective: Conventional cochlear implants provide patients who are deaf with hearing via electrical intracochlear stimulations. Stimulation electrodes are inserted into the cochlea through a cochleostomy or round window membrane (RWM) approach. However, these methods might induce cochlear ossificans and loss of residual hearing by damaging inner ear structures. To avoid an invasive electrode insertion, we developed a novel bone-guided extracochlear implant that stimulated the auditory nerves between the cochlear bones and the RWM to prevent cochlea damage. Power consumption plays an important role in wireless implantable electronic devices. Therefore, we aimed to investigate the effects of different electrodes on the stimulating threshold currents of the auditory nerve and the power consumption of bone-guided extracochlear implants using a commercial stimulator. Materials and Methods: Inert aurum (Au) electrodes were compared with biocompatible platinum (Pt) and iridium oxide (IrOx) electrodes in practical implantable applications. IrOxelectrodes were used for their high-charge storage capacity, low impedance, and biocompatibility. The electrodes were fabricated via sputtering and were experimentally characterized with cyclic voltammetry and then examined using in vivo tests. Results: Based on electrical auditory brainstem responses, IrOxelectrodes yielded lower acoustic nerve-stimulating threshold currents (132 μA) compared with Au electrodes (204 μA). IrOxelectrodes also had a lower acoustic nerve stimulating threshold current (132 μA) compared with Pt electrodes (168 μA). Conclusion: As expected, IrOxelectrodes were beneficial in the development of multielectrode bone-guided extracochlear implants, with the lowest acoustic nerve-stimulating threshold and current consumptions compared with Au and Pt electrodes.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed395    
    Printed25    
    Emailed0    
    PDF Downloaded30    
    Comments [Add]    

Recommend this journal